Back to Search Start Over

Intraocular Microsurgical Forceps (20, 23, and 25 gauge) Membrane Peeling Forces Assessment

Authors :
Raul Velez-Montoya
Chirag Patel
Scott C. N. Oliver
Hugo Quiroz-Mercado
Naresh Mandava
Jeffrey L. Olson
Source :
Journal of Ophthalmology, Vol 2013 (2013)
Publication Year :
2013
Publisher :
Hindawi Limited, 2013.

Abstract

Background. To assess the peeling forces exerted by different calibers of microsurgical forceps on an experimental model of epiretinal membrane. Methods. A model of epiretinal membrane was constructed using thin cellulose paper and heptanes-isopropyl alcohol 1% mixture. The model was mounted on a force censoring device. Subsequently, flaps were created with three different microsurgical forceps of different calibers. We recorded the number of attempts, the duration of the event, and the pushing and the pulling forces during the peeling. The results were compared by a one-way ANOVA and a Fisher unprotected least significant difference test with an alpha value of 0.05 for statistically significance. Results. There was a statistical significant difference on the pulling and pushing forces between the 25 gauge (13.79 mN; −13.27 mN) and the 23 (6.63 mN; −5.76 mN) and 20 (5.02 mN; −5.30 mN) gauge, being greater in the first (). There were no differences in the duration of all events, meaning that all the forces were measured within the same period of time. Conclusions. The 25 gauge microsurgical forceps exerted the greatest mechanical stress over our simulated epiretinal membrane model and required more attempts to create a surgical suitable flap. The clinical implication of this finding is still to be determined.

Subjects

Subjects :
Ophthalmology
RE1-994

Details

Language :
English
ISSN :
2090004X and 20900058
Volume :
2013
Database :
Directory of Open Access Journals
Journal :
Journal of Ophthalmology
Publication Type :
Academic Journal
Accession number :
edsdoj.600888e023949dd8a9802e26e78ee1b
Document Type :
article
Full Text :
https://doi.org/10.1155/2013/784172