In this article, MyFlex- $\epsilon $ , an ESR foot prosthesis equipped with a light and manually adjustable mechanism that allows for varying its stiffness in the sagittal plane, and a systematic approach to calculate its rotation-stiffness curves are presented. Through a design of experiment conducted numerically using a two-dimensional (2D) finite element (FE) model, calibrated experimentally, a geometric parameter whose variation alters the sagittal plane stiffness of a prosthesis originally designed with invariable stiffness, MyFlex- $\delta $ , was determined. After building the mechanism and integrating it into MyFlex- $\delta $ to obtain MyFlex- $\epsilon $ , the displacement-force curves of the latter through tests equivalent to the static tests specified in ISO 10328 were determined. Based on the experimental results, the 2D FE model of MyFlex- $\epsilon $ was built and calibrated to determine its rotation-stiffness curves in the sagittal plane. Comparing the rotation-stiffness curves obtained with the most compliant setting to the stiffest setting, stiffness variations of 119%, 122%, 138%, and 162% at plantarflexion angles of −5° and −2.5°, and dorsiflexion angles of 7.5° and 15°, respectively, were found.

Details

Language :
English
ISSN :
21693536
Volume :
12
Database :
Directory of Open Access Journals
Journal :
IEEE Access
Publication Type :
Academic Journal
Accession number :
edsdoj.6224212701434dc184540a9d76cffe27
Document Type :
article
Full Text :
https://doi.org/10.1109/ACCESS.2024.3427391