Back to Search Start Over

Unexpected Acceleration in Treprostinil Delivery Administered by a Lenus Pro® Implantable Pump in Two Patients Treated for Pulmonary Arterial Hypertension

Authors :
Garance Kopp
Anne-Lise Hachulla
Stéphane Noble
Aurélien Bringard
Paola M. Soccal
Maurice Beghetti
Frédéric Lador
Source :
Frontiers in Medicine, Vol 7 (2020)
Publication Year :
2020
Publisher :
Frontiers Media S.A., 2020.

Abstract

Intravenous treprostinil administration by an implantable pump is an attractive option for pulmonary arterial hypertension (PAH) treatment and is the subject of recent publications. Short-term studies are promising, but there is still a lack of long-term prospective data. We analyzed the treprostinil flow rate administered by the Lenus Pro® implantable pump in 2 patients suffering from PAH during follow-up times of respectively 4.2 and 3 years. The flow rate delivered by the pumps in these 2 patients exceeded the manufacturer admitted margin of error within 2 years and continued to increase to reach, respectively, 158 and 120% of the expected flow rate at the end of the follow up. In one case, the implantable pump had to be removed for this reason. The ex-vivo flow rate of the withdrawn pump determined in the laboratory reached 173% of the predicted value. This correlated with the in-vivo measurement, which suggests a continuous flow increase even after pump removal and without treprostinil use. Spontaneous flow increase from such an implantable pump is a potentially major pitfall, which needs to be identified and actively managed by the responsible clinicians.

Details

Language :
English
ISSN :
2296858X
Volume :
7
Database :
Directory of Open Access Journals
Journal :
Frontiers in Medicine
Publication Type :
Academic Journal
Accession number :
edsdoj.6454f07c8b2c4471ac95670e74ae1cf5
Document Type :
article
Full Text :
https://doi.org/10.3389/fmed.2020.539707