Back to Search Start Over

Risperdal® CONSTA® Needle Detachment. Incidence Rates Before and After Kit Redesign: A Retrospective Study using Electronic Health Records and Natural Language Processing in the Department of Veterans Affairs

Authors :
Marsha A. Wilcox
Danielle Coppola
Nicole Bailey
Andrew Wilson
Aaron W. C. Kamauu
Patrick R. Alba
Olga V. Patterson
Benjamin Viernes
Daniel W. Denhalter
Ira D. Solomon
Scott L. DuVall
Source :
Neurology and Therapy, Vol 8, Iss 1, Pp 95-108 (2019)
Publication Year :
2019
Publisher :
Adis, Springer Healthcare, 2019.

Abstract

Abstract Introduction Janssen received reports of needle detachments for Risperdal® CONSTA® and, in response, redesigned the kit. Objective The study objective was to estimate the rate of Risperdal® CONSTA® needle detachments prior to and after the introduction of a redesigned kit. Methods This retrospective study used record abstraction in the US Department of Veterans Affairs (VA). The 3 phases included: (1) a pilot study for methods evaluation in a sample of 6 hospitals with previously reported detachments; (2) a baseline study to ascertain the baseline detachment rate; and (3) a follow-up study to ascertain the rate for the redesigned kit. Administrative codes and natural language processing with clinical review were used to identify detachments. Results Pilot: we identified a subset of spontaneously reported detachments and several previously unreported events. In the baseline study (original device), from January through December 2013, 22 needle detachments were identified among 47,934 administrations of the drug in a census of administrations in the VA; an incidence of 0.0459%. In the follow-up study (redesigned device), from December 2015 through December 2016, there were 14 reported detachments in 41,819 injections, 0.0335%. This represents a reduction of 27% from the baseline. Conclusion This approach enabled us to identify needle detachments we would not have otherwise found (“solicited”). However, it likely resulted in incomplete outcome ascertainment. While this may have resulted in lower overall rates, it did not bias the comparison of the baseline and follow-up studies. The results showed that the redesigned Risperdal® CONSTA® kit reduced the incidence of needle detachment events in the VA. Funding Janssen Pharmaceuticals, Inc.

Details

Language :
English
ISSN :
21938253 and 21936536
Volume :
8
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Neurology and Therapy
Publication Type :
Academic Journal
Accession number :
edsdoj.65e87f8a25ff46749125fdfec58674b4
Document Type :
article
Full Text :
https://doi.org/10.1007/s40120-019-0130-7