Back to Search Start Over

Assessment of Open-path Spectrometer Accuracy at Low Path-integrated Methane Concentrations

Authors :
Zachary J. DeBruyn
Claudia Wagner-Riddle
Andrew VanderZaag
Source :
Atmosphere, Vol 11, Iss 2, p 184 (2020)
Publication Year :
2020
Publisher :
MDPI AG, 2020.

Abstract

The accurate measurement of greenhouse gas emissions is a challenge for atmospheric science. Long-range open-path sensors are flexible enough to be applied to a variety of complex emission sources, and single devices are often used to measure both high and low path-integrated concentrations. As this technology develops, it is important to examine potential sources of inaccuracy. A GasFinder3 open-path laser was tested with a range of path-integrated concentrations from 11.7 to 182 ppm∙m CH4 using certified standard gases. The measured path-integrated concentrations had a positive bias which was higher than 10% at low path-integrated concentrations (