Back to Search Start Over

Static vs. Dynamic Electrostatic Repulsion Reversed Phase Liquid Chromatography: Solutions for Pharmaceutical and Biopharmaceutical Basic Compounds

Authors :
Giulia Mazzoccanti
Francesco Gasparrini
Andrea Calcaterra
Claudio Villani
Alessia Ciogli
Source :
Separations, Vol 8, Iss 5, p 59 (2021)
Publication Year :
2021
Publisher :
MDPI AG, 2021.

Abstract

Many efforts have been made to separate basic compounds, which are challenging to resolve in reversed phase liquid chromatography. In this process, they are strongly retained and the peak shape undergoes significant distortion. The principal origin of this has been identified with the non-negligible interaction with residual deprotonated silanols. Consequently, all solutions that efficiently shield silanols are being sought. This review is an upgrade on the use of the electrostatic repulsion reversed phase (ERRP) approach: retention of bases, in protonated form, can be achieved by modulating the charge repulsion caused by the presence of positive charges in the chromatographic system. This study successfully (i) introduced fixed positive charges in the structure of stationary phases, (ii) used cationic and hydrophobic additives in the mobile phase, and (iii) used the ERRP-like approach employed at the preparative level for peptide purification.

Details

Language :
English
ISSN :
22978739
Volume :
8
Issue :
5
Database :
Directory of Open Access Journals
Journal :
Separations
Publication Type :
Academic Journal
Accession number :
edsdoj.75aff68f7c3848389ea138500b238315
Document Type :
article
Full Text :
https://doi.org/10.3390/separations8050059