Clinical Evaluation of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay

Authors :
Wai-Sing Chan
Kan-Pui Wong
Siu-Kei Yau
Ching-Yan Wong
Tsz-Ching Chan
Jeffrey Hung
Kristi Tsz-Wan Lai
Chin-Pang Leung
Candy Ling-Na Wang
Chun-Hang Au
Thomas Shek-Kong Wan
Edmond Shiu-Kwan Ma
Bone Siu-Fai Tang
Source :
Diagnostics, Vol 14, Iss 7, p 683 (2024)
Publication Year :
2024
Publisher :
MDPI AG, 2024.

Abstract

The performance of the Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assays were evaluated using 167 specimens, including 158 human respiratory specimens and 9 external quality assessment program (EQAP) samples. For respiratory specimens, CoV-2/Flu/RSV plus exhibited perfect agreement with the standard-of-care (SOC) methods (Cohen’s κ: 1, 100% agreement). The overall positive and negative percent agreement (PPA and NPA) were 100%, with 95% confidence intervals of 96.50 to 100% and 85.70 to 100%, respectively. On the other hand, Resp-4-Plex revealed an almost perfect agreement with the SOC methods (Cohen’s κ: 0.92, 97.71% agreement). The overall PPA and NPA were 100% (95.76 to 100%) and 88.46% (70.20 to 96.82%), respectively. For EQAP samples, the results of CoV-2/Flu/RSV plus (9/9) and Resp-4-Plex (4/4) were concordant with the expected results. The experimental limit of detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was the lowest (25 copies/mL for both methods), and that of the respiratory syncytial virus was the highest (400 copies/mL for CoV-2/Flu/RSV plus and 100 copies/mL for Resp-4-Plex). Threshold cycle (Ct) value correlation showed a large positive linear association between CoV-2/Flu/RSV plus and Resp-4-Plex, with R-squared values of 0.92–0.97, and on average, the Ct values of CoV-2/Flu/RSV plus were higher than that of Resp-4-Plex by 1.86–2.78, except for Flu A1 target (−0.66). To conclude, the performance of both assay was comparable to the SOC methods for both upper and lower respiratory specimens. Implementation of these rapid assay may reinforce the diagnostic capacity for the post-pandemic co-circulation of SARS-CoV-2 and other respiratory viruses.

Details

Language :
English
ISSN :
20754418
Volume :
14
Issue :
7
Database :
Directory of Open Access Journals
Journal :
Diagnostics
Publication Type :
Academic Journal
Accession number :
edsdoj.76ead1194b4e6586112c7a6ffad503
Document Type :
article
Full Text :
https://doi.org/10.3390/diagnostics14070683