Back to Search Start Over

A prospective, randomized comparison of the LMA-protector™ and i-gel™ in paralyzed, anesthetized patients

Authors :
Jee-Eun Chang
Hyerim Kim
Jung-Man Lee
Seong-Won Min
Dongwook Won
Kwanghoon Jun
Jin-Young Hwang
Source :
BMC Anesthesiology, Vol 19, Iss 1, Pp 1-7 (2019)
Publication Year :
2019
Publisher :
BMC, 2019.

Abstract

Abstract Background In the present study, we compare the LMA-Protector™ and the i-gel™ in terms of adequacy of the airway seal, insertion time, ease and accuracy of insertion, and the incidence of postoperative sore throat. Methods In 110 anesthetized and paralyzed adult patients, the i-gel™ (n = 55) or the LMA-Protector™ (n = 55) was inserted. The primary outcome was airway leak pressure. The secondary outcomes included the first-attempt success rate, insertion time, ease and accuracy of the device insertion, ease of gastric tube placement, blood staining on the device after removal, and incidence and severity of postoperative sore throat. Results The airway leak pressure was higher with the LMA-Protector™ than with the i-gel™ (31 [7] cmH2O vs. 27 [6] cmH2O, respectively; P = 0.016). Insertion time was longer with the LMA-Protector™ than with the i-gel™ (27 [16] sec vs. 19 [16] sec, respectively, P

Details

Language :
English
ISSN :
14712253
Volume :
19
Issue :
1
Database :
Directory of Open Access Journals
Journal :
BMC Anesthesiology
Publication Type :
Academic Journal
Accession number :
edsdoj.7f11e35d7a043758d83e028c89375de
Document Type :
article
Full Text :
https://doi.org/10.1186/s12871-019-0785-8