Differences in ultrasound elevational beam width (slice thickness) between popular handheld devices

Authors :
Harm J. Scholten
Gert Weijers
Marco de Wild
Hendrikus H.M. Korsten
Chris L. de Korte
R. Arthur Bouwman
Source :
WFUMB Ultrasound Open, Vol 1, Iss 2, Pp 100009- (2023)
Publication Year :
2023
Publisher :
Elsevier, 2023.

Abstract

Background: Handheld ultrasound devices are increasingly used by medical professionals for bedside ultrasound-guided interventions. Especially for vascular access procedures, the width of the imaging plane, known as the slice thickness or elevational beam width is a prominent source for misinterpretation. A wide slice thickness can lead to the interpretation that 2 objects (i.e. needle and vessel) are on the same plane while in fact they are not and thereby negatively influencing the performance of in-plane ultrasound-guided interventions. Therefore, the beam profiles of three popular handheld US devices are tested and compared to a conventional US device. Methods: The GE VScan, Philips Lumify and Butterfly IQ ​+ ​are tested using a slice phantom to determine the slice thickness. For comparison, a Philips Affiniti machine was investigated. Both linear and curved array settings were analyzed. In a slice phantom, a diffuse scattering plane at an angle of exactly 45° is scanned. For each imaging depth, the vertical height of the imaged rectangle corresponds to the slice thickness at that depth. Main results: For the linear array transducers, the focus depth ranges from 1.5 to 3.5 cm. At the focus depth, all transducers have a reasonable slice thickness of approximately 1 mm. More superficially, the slice thickness varies between 1 and 4 mm. The curved array probes have larger focus depths, ranging from 2.7 to 7.3 cm. The slice thickness at focus depth varies between 1.4 and 3.8 mm, but at 2 cm depth is even more than 5 mm. Conclusions: The slice thickness of handheld ultrasound transducers varies between the different devices, and can be suboptimal for superficial in-plane ultrasound-guided interventions. The larger slice thickness of the curved array settings may complicate in-plane guidance. Handheld ultrasound users should be aware of the beam characteristics of their devices to optimize guidance for interventions.

Details

Language :
English
ISSN :
29496683
Volume :
1
Issue :
2
Database :
Directory of Open Access Journals
Journal :
WFUMB Ultrasound Open
Publication Type :
Academic Journal
Accession number :
edsdoj.860ab522b4004da5a74ffc0cc7adb3a4
Document Type :
article
Full Text :
https://doi.org/10.1016/j.wfumbo.2023.100009