A modified open surgical revision technique for malfunctioning peritoneal dialysis catheter

Authors :
Shilong Xiang
Xiaohui Zhang
Yaomin Wang
Guangjun Liu
Xishao Xie
Fei Han
Jianghua Chen
Source :
Renal Failure, Vol 46, Iss 2 (2024)
Publication Year :
2024
Publisher :
Taylor & Francis Group, 2024.

Abstract

Backgrounds: The malfunction of peritoneal dialysis (PD) catheter is still an intractable problem. A modified open surgical revision technique with suturing fixation and without catheter removal for malfunctioning catheter was developed to evaluated the efficacy and safety between simultaneous catheter replacement technique.Methods: A total of 167 PD patients with malfunctioning catheter were retrospectively reviewed. For the salvage of PD catheters, patients underwent modified open surgical revision (group A) or simultaneous catheter replacement (group B). The baseline characteristics before operation, perioperative condition, complications and outcomes were compared between both groups.Results: Patients of group A showed significantly shorter operative time (67.4 ± 22.1 versus 82.8 ± 21.1 min, p = 0.009), less postoperative pain score within 24 h (median 0.0 versus 2.0, p