Cone beam CT for perioperative imaging in hearing preservation Cochlear implantation – a human cadaveric study

Authors :
Kayvan Nateghifard
David Low
Lola Awofala
Dilakshan Srikanthan
Jafri Kuthubutheen
Michael Daly
Harley Chan
Jonathan Irish
Joseph Chen
Vincent Lin
Trung Ngoc Le
Source :
Journal of Otolaryngology - Head and Neck Surgery, Vol 48, Iss 1, Pp 1-8 (2019)
Publication Year :
2019
Publisher :
BMC, 2019.

Abstract

Abstract Background Knowledge of the cochlear implant array’s precise position is important because of the correlation between electrode position and speech understanding. Several groups have provided recent image processing evidence to determine scalar translocation, angular insertion depth, and cochlear duct length (CDL); all of which are being used for patient-specific programming. Cone beam computed tomography (CBCT) is increasingly used in otology due to its superior resolution and low radiation dose. Our objectives are as followed: 1.Validate CBCT by measuring cochlear metrics, including basal turn diameter (A-value) and lateral wall cochlear duct length at different angular intervals and comparing it against microcomputed CT (uCT).2.Explore the relationship between measured lateral wall cochlear duct length at different angular intervals and insertion depth among 3 different length electrodes using CBCT. Methods The study was performed using fixed human cadaveric temporal bones in a tertiary academic centre. Ten temporal bones were subjected to the standard facial recess approach for cochlear implantation and imaged by CBCT followed by uCT. Measurements were performed on a three-dimensional reconstructed model of the cochlea. Sequential insertion of 3 electrodes (Med-El Flex24, 28 and Soft) was then performed in 5 bones and reimaged by CBCT. Statistical analysis was performed using Pearson’s correlation. Results There was good agreement between CBCT and uCT for cochlear metrics, validating the precision of CBCT against the current gold standard uCT in imaging. The A-value recorded by both modalities showed a high degree of linear correlation and did not differ by more than 0.23 mm in absolute values. For the measurement of lateral wall CDL at various points along the cochlea, there was a good correlation between both modalities at 360 deg and 720 deg (r = 0.85, p

Subjects

Subjects :
Surgery
RD1-811

Details

Language :
English
ISSN :
19160216
Volume :
48
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Journal of Otolaryngology - Head and Neck Surgery
Publication Type :
Academic Journal
Accession number :
edsdoj.88b41d0e90314643a3612eeb2ea415b6
Document Type :
article
Full Text :
https://doi.org/10.1186/s40463-019-0388-x