Impact of iodine concentration and scan parameters on image quality, contrast enhancement and radiation dose in thoracic CT

Authors :
Marian S. Solbak
Mette K. Henning
Andrew England
Anne C. Martinsen
Trond M. Aaløkken
Safora Johansen
Source :
European Radiology Experimental, Vol 4, Iss 1, Pp 1-8 (2020)
Publication Year :
2020
Publisher :
SpringerOpen, 2020.

Abstract

Abstract Background We investigated the impact of varying contrast medium (CM) densities and x-ray tube potentials on contrast enhancement (CE), image quality and radiation dose in thoracic computed tomography (CT) using two different scanning techniques. Methods Seven plastic tubes containing seven different CM densities ranging from of 0 to 600 HU were positioned inside a commercial chest phantom with padding, representing three different patient sizes. Helical scans of the phantom in single-source mode were obtained with varying tube potentials from 70 to 140 kVp. A constant volume CT dose index (CTDIvol) depending on phantom size and automatic dose modulation was tested. CE (HU) and image quality (contrast-to-noise ratio, CNR) were measured for all combinations of CM density and tube potential. A reference threshold of CE and kVp was defined as ≥ 200 HU and 120 kVp. Results For the medium-sized phantom, with a specific CE of 100–600 HU, the diagnostic CE (200 HU) at 70 kVp was ~ 90% higher than at 120 kVp, for both scan techniques (p < 0.001). Changes in CM density/specific HU together with lower kVp resulted in significantly higher CE and CNR (p < 0.001). When changing only the kVp, no statistically significant differences were observed in CE or CNR (p ≥ 0.094), using both dose modulation and constant CTDIvol. Conclusions For thoracic CT, diagnostic CE (≥ 200 HU) and maintained CNR were achieved by using lower CM density in combination with lower tube potential (< 120 kVp), independently of phantom size.

Details

Language :
English
ISSN :
25099280
Volume :
4
Issue :
1
Database :
Directory of Open Access Journals
Journal :
European Radiology Experimental
Publication Type :
Academic Journal
Accession number :
edsdoj.8d41182ff90f4a0e8653fc784fde316c
Document Type :
article
Full Text :
https://doi.org/10.1186/s41747-020-00184-z