Back to Search Start Over

A Concept for a Sensitive Micro Total Analysis System for High Throughput Fluorescence Imaging

Authors :
Yosi Shacham
Arthur Rabner
Source :
Sensors, Vol 6, Iss 4, Pp 341-349 (2006)
Publication Year :
2006
Publisher :
MDPI AG, 2006.

Abstract

This paper discusses possible methods for on-chip fluorescent imaging forintegrated bio-sensors. The integration of optical and electro-optical accessories, accordingto suggested methods, can improve the performance of fluorescence imaging. It can boostthe signal to background ratio by a few orders of magnitudes in comparison to conventionaldiscrete setups. The methods that are present in this paper are oriented towards buildingreproducible arrays for high-throughput micro total analysis systems (μTAS). The firstmethod relates to side illumination of the fluorescent material placed into micro-compartments of the lab-on-chip. Its significance is in high utilization of excitation energyfor low concentration of fluorescent material. The utilization of a transparent μLED chip,for the second method, allows the placement of the excitation light sources on the sameoptical axis with emission detector, such that the excitation and emission rays are directedcontroversly. The third method presents a spatial filtering of the excitation background.

Details

Language :
English
ISSN :
14248220
Volume :
6
Issue :
4
Database :
Directory of Open Access Journals
Journal :
Sensors
Publication Type :
Academic Journal
Accession number :
edsdoj.8e90b7b6b9e245489f5fa04dec2208e4
Document Type :
article
Full Text :
https://doi.org/10.3390/S6040341