Back to Search Start Over

Smartphone System for Heart Rate and Breathing Rate Estimation

Authors :
Amit Nayak
Miodrag Bolic
Source :
IEEE Open Journal of Instrumentation and Measurement, Vol 3, Pp 1-4 (2024)
Publication Year :
2024
Publisher :
IEEE, 2024.

Abstract

In this short article, we present a new method to use a smartphone placed unattached on a subject’s chest in the supine position to obtain heartbeat and breathing signals and estimate heart and breathing rates, simultaneously. We collected 3-axis accelerometer, gyroscope, and magnetometer signals and performed sensor fusion to extract a user’s breathing signal and breathing rate. A hidden Markov model was used to segment the ballistocardiograph/seismocardiograph signals and extract the heart rate. The smartphone application was verified against breathing belt measurements and electrocardiogram measurements. We modified and proposed several suitable signal quality metrics for seismocardiograph signals. The overall results show that the application accurately estimated the breathing and heart rates, achieving a minimum mean percent error of 2.52% for breathing and 2.33% for heart rate. This work is a big step forward for vital sign estimation using inexpensive pervasive devices.

Details

Language :
English
ISSN :
27687236
Volume :
3
Database :
Directory of Open Access Journals
Journal :
IEEE Open Journal of Instrumentation and Measurement
Publication Type :
Academic Journal
Accession number :
edsdoj.94d60317e3494086ad7a91420b8d5bba
Document Type :
article
Full Text :
https://doi.org/10.1109/OJIM.2024.3477572