Evaluation of repeatability and agreement of two optical biometers for intraocular lens power calculation

Authors :
Irene Martinez Alberquilla
Saga Svensson
Javier Ruiz-Alcocer
David Madrid-Costa
Alberto Dominguez-Vicent
Abinaya Priya Venkataraman
Source :
Scientific Reports, Vol 14, Iss 1, Pp 1-8 (2024)
Publication Year :
2024
Publisher :
Nature Portfolio, 2024.

Abstract

Abstract The repeatability of two biometers (Lenstar-LS900 and Eyestar-900) to measure ocular parameters and intraocular lens (IOL) power calculation, and their agreement were evaluated. 134 eyes of 134 participants were measured thrice with each biometer. Axial length (AL), anterior chamber depth (ACD), lens thickness (LT) and keratometry (K) were evaluated. The IOL power was calculated using different formulas. The repeatability limit (RLimit), the mean differences (MD) and the limits of agreement (LoA) were calculated. The RLimits for all parameters were higher with Lenstar compared to Eyestar. RLimits were lower than 0.50 D except for Barrett Universal II (0.54 D) and Haigis (0.51 D) formulas with the Lenstar. Mean differences were lower than 0.01 mm for AL, ACD and LT, and lower than 0.03 D for K. MD ranged from 0 to 0.02 D for all formulas except for Barrett and Hill. When dividing the sample into subgroups (short, normal and long eyes), the MDs were similar for the IOL power and were lower than 0.03 D, except for the Barrett and Hill formulas. Both biometers provide repeatable biometry and IOL power calculations. The LoA interval for the IOL power calculation was between 0.75 and 1.50D, which was similar among the subgroups.

Details

Language :
English
ISSN :
20452322
Volume :
14
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Scientific Reports
Publication Type :
Academic Journal
Accession number :
edsdoj.9fc43d5bb7cd4177ad405cb37b174354
Document Type :
article
Full Text :
https://doi.org/10.1038/s41598-024-73206-0