Back to Search Start Over

In Vivo Evaluation of Glucose Permeability of an Immunoisolation Device Intended for Islet Transplantation: A Novel Application of the Microdialysis Technique

Authors :
E. Rafael
A. Wernerson
P. Arner
G. S. Wu
A. Tibell
Source :
Cell Transplantation, Vol 8 (1999)
Publication Year :
1999
Publisher :
SAGE Publishing, 1999.

Abstract

Immunoisolation devices consist of semipermeable membranes chosen to protect the islets from the immune system but still allow sufficient passage of nutrients, oxygen, and the therapeutic products, insulin. The exchange between the device and the microcirculation will influence the survival of the graft as well as the metabolic efficacy of the islet implant. Glucose is the important trigger factor for insulin secretion. In this study, we evaluate the in vivo glucose permeability of the Theracyte™ immunoisolation device at various times after implantation. Empty devices were implanted SC in rats. The glucose kinetics in the device was compared to that in the SC tissue during IV glucose tolerance tests (IVGTTs), using the microdialysis technique. In rats studied on day 1, or 1, 2, and 4 weeks after implantation, the peak glucose levels (C max ) were significantly lower, the times-to-peak (TTP) were significantly longer, and the areas under the curve during the first 40 min (AUC 0–40 ) were significantly smaller in the device than in the SC fat. However, at 3 months all parameters improved and C max , TTP, and AUC 0–40 in the device did not differ significantly from those measured in the SC fat. Thus, during the first 4 weeks the device constitutes a significant diffusion barrier, but at 3 months the exchange between the lumen of devices and the blood stream improves. Our data indicate that implantation of the device several months before transplantation of the cellular graft would improve the exchange across the membrane during the early posttransplant period. This should have positive effects on graft survival and function. We also suggest that microdialysis is a useful tool for evaluating the in vivo performance of macroencapsulation devices.

Subjects

Subjects :
Medicine

Details

Language :
English
ISSN :
09636897 and 15553892
Volume :
8
Database :
Directory of Open Access Journals
Journal :
Cell Transplantation
Publication Type :
Academic Journal
Accession number :
edsdoj.b1366a9e6864a6f88fe1bbd2c963399
Document Type :
article
Full Text :
https://doi.org/10.1177/096368979900800302