Design, fabrication, and validation of a polymethyl methacrylate head phantom for dosimetric verification of cranial radiotherapy treatment plans

Authors :
V S Shaiju
Rajesh Kumar
Debjani Phani
K V Rajasekhar
George Zacharia
Saju Bhasi
Raghuram K Nair
Source :
Journal of Medical Physics, Vol 45, Iss 2, Pp 66-70 (2020)
Publication Year :
2020
Publisher :
Wolters Kluwer Medknow Publications, 2020.

Abstract

Purpose: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. Materials and Methods: The head phantom designing involves slice-wise modeling of an adult head using PMMA. The phantom has provisions to hold detectors such as ionization chambers of different sizes, Gafchromic films, gel dosimeter, and optically stimulated luminescence dosimeter. For the point dose verification purpose, 15 volumetric modulated arc therapy patient plans were selected, and doses were measured using a CC13 ionization chamber. The percentage gamma passing rate was calculated for acceptance criteria 3%/3 mm and 2%/2 mm using OmniPro I'mRT film QA software, and Gafchromic EBT3 films were used for 2D planar dose verification. Results: Treatment planning system calculated, and the measured point doses showed a percentage deviation ranged from 0.26 to 1.92. The planar dose fluence measurements, for set acceptance criteria of 3%/3 mm and 2%/2 mm, percentages of points having gamma value

Details

Language :
English
ISSN :
09716203 and 19983913
Volume :
45
Issue :
2
Database :
Directory of Open Access Journals
Journal :
Journal of Medical Physics
Publication Type :
Academic Journal
Accession number :
edsdoj.b677804e2b8f41a3ae3f1fdfe02e7563
Document Type :
article
Full Text :
https://doi.org/10.4103/jmp.JMP_21_20