Back to Search Start Over

A lightweight magnetically shielded room with active shielding

Authors :
Niall Holmes
Molly Rea
James Chalmers
James Leggett
Lucy J. Edwards
Paul Nell
Stephen Pink
Prashant Patel
Jack Wood
Nick Murby
David Woolger
Eliot Dawson
Christopher Mariani
Tim M. Tierney
Stephanie Mellor
George C. O’Neill
Elena Boto
Ryan M. Hill
Vishal Shah
James Osborne
Rosemarie Pardington
Peter Fierlinger
Gareth R. Barnes
Paul Glover
Matthew J. Brookes
Richard Bowtell
Source :
Scientific Reports, Vol 12, Iss 1, Pp 1-13 (2022)
Publication Year :
2022
Publisher :
Nature Portfolio, 2022.

Abstract

Abstract Magnetically shielded rooms (MSRs) use multiple layers of materials such as MuMetal to screen external magnetic fields that would otherwise interfere with high precision magnetic field measurements such as magnetoencephalography (MEG). Optically pumped magnetometers (OPMs) have enabled the development of wearable MEG systems which have the potential to provide a motion tolerant functional brain imaging system with high spatiotemporal resolution. Despite significant promise, OPMs impose stringent magnetic shielding requirements, operating around a zero magnetic field resonance within a dynamic range of ± 5 nT. MSRs developed for OPM-MEG must therefore effectively shield external sources and provide a low remnant magnetic field inside the enclosure. Existing MSRs optimised for OPM-MEG are expensive, heavy, and difficult to site. Electromagnetic coils are used to further cancel the remnant field inside the MSR enabling participant movements during OPM-MEG, but present coil systems are challenging to engineer and occupy space in the MSR limiting participant movements and negatively impacting patient experience. Here we present a lightweight MSR design (30% reduction in weight and 40–60% reduction in external dimensions compared to a standard OPM-optimised MSR) which takes significant steps towards addressing these barriers. We also designed a ‘window coil’ active shielding system, featuring a series of simple rectangular coils placed directly onto the walls of the MSR. By mapping the remnant magnetic field inside the MSR, and the magnetic field produced by the coils, we can identify optimal coil currents and cancel the remnant magnetic field over the central cubic metre to just |B|= 670 ± 160 pT. These advances reduce the cost, installation time and siting restrictions of MSRs which will be essential for the widespread deployment of OPM-MEG.

Subjects

Subjects :
Medicine
Science

Details

Language :
English
ISSN :
20452322
Volume :
12
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Scientific Reports
Publication Type :
Academic Journal
Accession number :
edsdoj.f1fffe3ca8934e74b2f35ead611cf2e2
Document Type :
article
Full Text :
https://doi.org/10.1038/s41598-022-17346-1