Back to Search Start Over

Efficacy of smartphone-based retinal photography by undergraduate students in screening and early diagnosing diabetic retinopathy

Authors :
Jéssica Deponti Gobbi
João Pedro Romero Braga
Moises M. Lucena
Victor C. F. Bellanda
Miguel V. S. Frasson
Daniel Ferraz
Victor Koh
Rodrigo Jorge
Source :
International Journal of Retina and Vitreous, Vol 8, Iss 1, Pp 1-9 (2022)
Publication Year :
2022
Publisher :
BMC, 2022.

Abstract

Abstract Background To evaluate the efficacy of retinal photography obtained by undergraduate students using a smartphone-based device in screening and early diagnosing diabetic retinopathy (DR). Methods We carried out an open prospective study with ninety-nine diabetic patients (194 eyes), who were submitted to an ophthalmological examination in which undergraduate students registered images of the fundus using a smartphone-based device. At the same occasion, an experienced nurse captured fundus photographs from the same patients using a gold standard tabletop camera system (Canon CR-2 Digital Non-Mydriatic Retinal Camera), with a 45º field of view. Two distinct masked specialists evaluated both forms of imaging according to the presence or absence of sings of DR and its markers of severity. We later compared those reports to assess agreement between the two technologies. Results Concerning the presence or absence of DR, we found an agreement rate of 84.07% between reports obtained from images of the smartphone-based device and from the regular (tabletop) fundus camera; Kappa: 0.67; Sensitivity: 71.0% (Confidence Interval [CI]: 65.05–78.16%); Specificity: 94.06% (CI: 90.63–97.49%); Accuracy: 84.07%; Positive Predictive Value (PPV): 90.62%; Negative Predictive Value (NPV): 80.51%. As for the classification between proliferative diabetic retinopathy and non-proliferative diabetic retinopathy, we found an agreement of 90.00% between the reports; Kappa: 0.78; Sensitivity: 86.96%; (CI: 79.07–94.85%); Specificity: 91.49% (CI: 84.95–98.03%); Accuracy: 90.00%; PPV: 83.33%; NPV: 93.48%. Regarding the degree of classification of DR, we found an agreement rate of 69.23% between the reports; Kappa: 0.52. As relating to the presence or absence of hard macular exudates, we found an agreement of 84.07% between the reports; Kappa: 0.67; Sensitivity: 71.60% (CI: 65.05–78.16%); Specificity: 94.06% (CI: 90.63–97.49%); Accuracy: 84.07%; PPV: 90.62%; NPV: 80.51%. Conclusion The smartphone-based device showed promising accuracy in the detection of DR (84.07%), making it a potential tool in the screening and early diagnosis of DR.

Details

Language :
English
ISSN :
20569920
Volume :
8
Issue :
1
Database :
Directory of Open Access Journals
Journal :
International Journal of Retina and Vitreous
Publication Type :
Academic Journal
Accession number :
edsdoj.f89a055bb9e3441faa3e179a8a7abcb3
Document Type :
article
Full Text :
https://doi.org/10.1186/s40942-022-00388-y