Back to Search
Start Over
Thermometry for turbulent flames by coherent anti-Stokes Raman spectroscopy with simultaneous referencing to the modeless excitation profile
- Source :
- Applied Optics. Nov 10, 2005, Vol. 44 Issue 32, p6995, 10 p.
- Publication Year :
- 2005
-
Abstract
- An optimal system for temperature measurements by coherent anti-Stokes Raman spectroscopy (CARS) in turbulent flames and flows is presented. In addition to a single-mode pump laser and a modeless dye laser, an echelle spectrometer with a cross disperser is used. This system permits simultaneous measurement of the [N.sub.2] CARS spectrum and the broadband dye laser profile. A procedure is developed to use software to transform this profile into the excitation profile by which the spectrum is referenced. Simultaneous shot-to-shot referencing is compared to sequential averaged referencing for data obtained in flat flames and in room air. At flame temperatures, the resultant 1.5% imprecision is limited by flame fluctuations, indicating that the system may have a single-shot imprecision below 1%. At room temperature, the 3.8% single-shot imprecision is of the same order as the best values reported for dualbroadband pure-rotational CARS. Using the unique shot-to-shot excitation profiles, simultaneous referencing eliminates systematic errors. At 2000 and 300 K, the 95% confidence intervals are estimated to be [+ or -]20 and [+ or -]10 K, respectively. OCIS codes: 120.1740, 120.6780, 190.1900, 300.6230.
Details
- Language :
- English
- ISSN :
- 1559128X
- Volume :
- 44
- Issue :
- 32
- Database :
- Gale General OneFile
- Journal :
- Applied Optics
- Publication Type :
- Academic Journal
- Accession number :
- edsgcl.139037018