Back to Search Start Over

Calibration for cylindrical specimens in grazing-incidence interferometry via integration of difference measurements

Authors :
Mantel, Klaus
Lamprecht, Jurgen
Lindlein, Norbert
Schwider, Johannes
Source :
Applied Optics. Nov 1, 2006, Vol. 45 Issue 31, p8013, 6 p.
Publication Year :
2006

Abstract

Cylindrical specimens may be tested advantageously by using grazing-incidence interferometry. A multiple positions test in combination with rotational averaging has recently been used to separate the surface deviations of the specimen from the interferometric aberrations. To reduce the measuring time and to check whether the results are reliable, a second procedure is now investigated, which uses the principle of the multiple positions test to determine quantities proportional to the difference quotients of the surface deviations. After numerical integration, the results can be compared with those obtained previously by rotational averaging. The measurement principle is described, and calibration results are presented.

Details

Language :
English
ISSN :
1559128X
Volume :
45
Issue :
31
Database :
Gale General OneFile
Journal :
Applied Optics
Publication Type :
Academic Journal
Accession number :
edsgcl.154067722