Position sensitivity of segmented planar HPGe detectors for the DESPEC project at FAIR
- Publication Year :
- 2008
-
Abstract
- One of the proposed germanium detector systems for the upcoming DESPEC array at the FAIR facility consists of triple modules of electrically segmented planar high-purity germanium detectors. In this work, the position sensitivity obtained by means of pulse shape analysis (PSA) for simulated gamma-ray interactions has been studied for the possible segmentation patterns for such detectors-the double-sided strip detector (DSSD) and the one-sided pixelated geometry. An over-all similar number of readout channels has been considered for each case, resulting in 8 + 8 strips and 16 pixels. It has been found that the higher physical granularity of the DSSD results in a significantly higher position resolution, as well as in a somewhat lower probability of merging multiple interaction points. A drawback of the DSSD geometry is the inability to assign the interactions to the correct voxels for certain complex multiple interaction patterns. This effect has been studied in detail. Finally, a comparison has been made with the existing 25-pixel planar detector at KTH [A. Khaplanov, et al., Nucl. Instr. and Meth. A 580 (2007) 1075; L. Milechina, B. Cederwall, Nucl. Instr. and Meth. A 550 (2005) 278)].<br />QC 20101110