Back to Search Start Over

Constructing carbon fiber motion-detection loops for simultaneous EEG-fMRI

Authors :
Abbott, DE
Masterton, RAJ
Archer, JS
Fleming, SW
Warren, AEL
Jackson, GD
Abbott, DE
Masterton, RAJ
Archer, JS
Fleming, SW
Warren, AEL
Jackson, GD
Publication Year :
2015

Abstract

One of the most significant impediments to high-quality EEG recorded in an MRI scanner is subject motion. Availability of motion artifact sensors can substantially improve the quality of the recorded EEG. In the study of epilepsy, it can also dramatically increase the confidence that one has in discriminating true epileptiform activity from artifact. This is due both to the reduction in artifact and the ability to visually inspect the motion sensor signals when reading the EEG, revealing whether or not head motion is present. We have previously described the use of carbon fiber loops for detecting and correcting artifact in EEG acquired simultaneously with MRI. The loops, attached to the subject's head, are electrically insulated from the scalp. They provide a simple and direct measure of specific artifact that is contaminating the EEG, including both subject motion and residual artifact arising from magnetic field gradients applied during MRI. Our previous implementation was used together with a custom-built EEG-fMRI system that differs substantially from current commercially available EEG-fMRI systems. The present technical note extends this work, describing in more detail how to construct the carbon fiber motion-detection loops, and how to interface them with a commercially available simultaneous EEG-fMRI system. We hope that the information provided may help those wishing to utilize a motion-detection/correction solution to improve the quality of EEG recorded within an MRI scanner.

Details

Database :
OAIster
Publication Type :
Electronic Resource
Accession number :
edsoai.on1315691769
Document Type :
Electronic Resource