Back to Search Start Over

A Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis: Interim Clinical Trial Results

Authors :
Petoe, MA
Titchener, SA
Kolic, M
Kentler, WG
Abbott, CJ
Nayagam, DAX
Baglin, EK
Kvansakul, J
Barnes, N
Walker, JG
Epp, SB
Young, KA
Ayton, LN
Luu, CD
Allen, PJ
Petoe, MA
Titchener, SA
Kolic, M
Kentler, WG
Abbott, CJ
Nayagam, DAX
Baglin, EK
Kvansakul, J
Barnes, N
Walker, JG
Epp, SB
Young, KA
Ayton, LN
Luu, CD
Allen, PJ
Publication Year :
2021

Abstract

PURPOSE: To report the initial safety and efficacy results of a second-generation (44-channel) suprachoroidal retinal prosthesis at 56 weeks after device activation. METHODS: Four subjects, with advanced retinitis pigmentosa and bare-light perception only, enrolled in a phase II trial (NCT03406416). A 44-channel electrode array was implanted in a suprachoroidal pocket. Device stability, efficacy, and adverse events were investigated at 12-week intervals. RESULTS: All four subjects were implanted successfully and there were no device-related serious adverse events. Color fundus photography indicated a mild postoperative subretinal hemorrhage in two recipients, which cleared spontaneously within 2 weeks. Optical coherence tomography confirmed device stability and position under the macula. Screen-based localization accuracy was significantly better for all subjects with device on versus device off. Two subjects were significantly better with the device on in a motion discrimination task at 7, 15, and 30°/s and in a spatial discrimination task at 0.033 cycles per degree. All subjects were more accurate with the device on than device off at walking toward a target on a modified door task, localizing and touching tabletop objects, and detecting obstacles in an obstacle avoidance task. A positive effect of the implant on subjects' daily lives was confirmed by an orientation and mobility assessor and subject self-report. CONCLUSIONS: These interim study data demonstrate that the suprachoroidal prosthesis is safe and provides significant improvements in functional vision, activities of daily living, and observer-rated quality of life. TRANSLATIONAL RELEVANCE: A suprachoroidal prosthesis can provide clinically useful artificial vision while maintaining a safe surgical profile.

Details

Database :
OAIster
Publication Type :
Electronic Resource
Accession number :
edsoai.on1315735400
Document Type :
Electronic Resource