Underfilled blood tube containing EDTA: Is it an inappropriate sample for HbA1c assay?

Authors :
Guven, Berrak
Benice, Ismail
Can, Murat
Guven, Berrak
Benice, Ismail
Can, Murat
Source :
Biochemia Medica; ISSN 1330-0962 (Print); ISSN 1846-7482 (Online); Volume 33; Issue 1
Publication Year :
2023

Abstract

IntroductionBlood samples having inappropriate volume are a substantial part of preanalytical errors. Inadequate sample volume for glycated haemoglobin (HbA1c) test may be a common problem of patients with diabetes mellitus having vascular changes. In this study, we compared HbA1c concentrations of underfilled and appropriately filled blood collection tubes. Materials and methodsTo compare HbA1c concentrations, blood samples were collected into 2 mL tubes containing K3-EDTA from 109 subjects. Two blood samples (underfilled and appropriately filled) were drawn from a patient by the same personnel and materials. HbA1c measurements were assayed on a Cobas 6000 analyser module c 501 (Roche Diagnostics, Mannheim, Germany). The HbA1c% results were compared by t-test and Wilcoxon’s signed-rank statistical methods (SPSS Inc., Chicago, USA). Bias analysis was performed using Microsoft Excel 4.0. ResultsUnderfilled samples were classified three groups (group 1, N = 44; group 2, N = 36; and group 3, N = 29) according to the filling ratio of the samples; 0.5 mL and below (< 25%), 0.5-1.0 mL (25-50%), and 1.0-2.0 mL (> 50%), respectively. When we compared underfilled tubes with pairing filled tubes, there was a statistically significant difference only with tubes filled less than 25% (P = 0.030). Furthermore, we have done bias analysis between paired tubes according to the diagnostic cut-off value of 6.5%. The bias was more prominent in up to 50% underfilled blood tubes (1.1%), when HbA1c concentrations were below the diagnostic cut-off of 6.5%. ConclusionsWe suggest that the blood tubes with EDTA for HbA1c measurement should be filled with at least 50% to avoid clinical variations.

Details

Database :
OAIster
Journal :
Biochemia Medica; ISSN 1330-0962 (Print); ISSN 1846-7482 (Online); Volume 33; Issue 1
Notes :
application/pdf, English
Publication Type :
Electronic Resource
Accession number :
edsoai.on1371042655
Document Type :
Electronic Resource