Back to Search Start Over

Comparison of Disk Diffusion, E-Test, and Broth Microdilution Methods for Testing In Vitro Activity of Cefiderocol in Acinetobacter baumannii

Authors :
Kolesnik-Goldmann, Natalia
Seth-Smith, Helena M B
Haldimann, Klara
Imkamp, Frank
Roloff, Tim; https://orcid.org/0000-0003-3435-4723
Zbinden, Reinhard
Hobbie, Sven N; https://orcid.org/0000-0002-9205-0384
Egli, Adrian; https://orcid.org/0000-0002-3564-8603
Mancini, Stefano; https://orcid.org/0000-0003-4516-5952
Kolesnik-Goldmann, Natalia
Seth-Smith, Helena M B
Haldimann, Klara
Imkamp, Frank
Roloff, Tim; https://orcid.org/0000-0003-3435-4723
Zbinden, Reinhard
Hobbie, Sven N; https://orcid.org/0000-0002-9205-0384
Egli, Adrian; https://orcid.org/0000-0002-3564-8603
Mancini, Stefano; https://orcid.org/0000-0003-4516-5952
Source :
Kolesnik-Goldmann, Natalia; Seth-Smith, Helena M B; Haldimann, Klara; Imkamp, Frank; Roloff, Tim; Zbinden, Reinhard; Hobbie, Sven N; Egli, Adrian; Mancini, Stefano (2023). Comparison of Disk Diffusion, E-Test, and Broth Microdilution Methods for Testing In Vitro Activity of Cefiderocol in Acinetobacter baumannii. Antibiotics, 12(7):1212.
Publication Year :
2023

Abstract

The reference method for cefiderocol antimicrobial susceptibility testing is broth microdilution (BMD) with iron-depleted-Mueller-Hinton (ID-MH) medium, whereas breakpoints recommended for disk diffusion (DD) are based on MH-agar plates. We aimed to compare the performance of the commercial BMD tests ComASP (Liofilchem) and UMIC (Bruker), and DD and E-test using MH- and ID-MH-agar plates with the reference BMD method using 100 carbapenem-resistant-A. baumannii isolates. Standard BMD was performed according to the EUCAST guidelines; DD and E-test were carried out using two commercial MH-agar plates (BioMérieux and Liofilchem) and an in-house ID-MH-agar plate, while ComASP and UMIC were performed according to the manufacturer's guidelines. DD performed with the ID-MH-agar plates led to a higher categorical agreement (CA, 95.1%) with standard BMD and fewer categorization errors compared to the commercial MH-agar plates (CA BioMérieux 91.1%, Liofilchem 89.2%). E-test on ID-MH-agar plates exhibited a significantly higher essential agreement (EA, 75%) with standard BMD compared to the two MH-agar plates (EA BioMérieux 57%, Liofilchem 44%), and showed a higher performance in detecting high-level resistance than ComASP and UMIC (mean log2 difference with standard BMD for resistant isolates of 0.5, 2.83, and 2.08, respectively). In conclusion, DD and E-test on ID-MH-agar plates exhibit a higher diagnostic performance than on MH-agar plates and the commercial BMD methods. Therefore, we recommend using ID-MH-agar plates for cefiderocol susceptibility testing of A. baumannii.

Details

Database :
OAIster
Journal :
Kolesnik-Goldmann, Natalia; Seth-Smith, Helena M B; Haldimann, Klara; Imkamp, Frank; Roloff, Tim; Zbinden, Reinhard; Hobbie, Sven N; Egli, Adrian; Mancini, Stefano (2023). Comparison of Disk Diffusion, E-Test, and Broth Microdilution Methods for Testing In Vitro Activity of Cefiderocol in Acinetobacter baumannii. Antibiotics, 12(7):1212.
Notes :
application/pdf, info:doi/10.5167/uzh-239652, English, English
Publication Type :
Electronic Resource
Accession number :
edsoai.on1443054815
Document Type :
Electronic Resource