1. Patent blue interferes with the measurement of lipemia index in a patient with sentinel lymph node.
- Author
-
Nakano, Keiichi, Seimiya, Masanori, Yamazaki, Kojiro, Yasuda, Keiko, Yamashita, Naoki, Goto, Hideki, and Teshima, Takanori
- Subjects
BLOOD testing ,LIPID analysis ,SENTINEL lymph node biopsy ,PHOTON absorptiometry ,COLORIMETRY ,SENTINEL lymph nodes ,BILIRUBIN ,DYES & dyeing ,URINALYSIS ,BASAL cell carcinoma - Abstract
Lipids interfere with absorbance measurements conducted using colorimetric methods. To monitor lipemia, some systems measure absorbance using an analyzer. This report describes a novel case of interference with the lipemia index without lipemia. A 64-year-old woman with giant basal cell carcinoma underwent resection and sentinel lymph node biopsy. The patient had been subcutaneously injected with patent blue during sentinel lymph node resection. After surgery, her serum and urine were yellow-green, and the lipemia index, calculated by measuring absorbance at 658 nm (main wavelength) and 694 nm (secondary wavelength) using a JCA-BM8040 chemistry analyzer, was high. The absorbance spectrum of the patient's serum and patent blue solution were compared to determine the cause of the high lipemia index. The patient's serum and the patent blue solution showed absorption at wavelengths between 540 and 698 nm. Moreover, the absorbance was concentration-dependent for patent blue. These results thus indicated that the patient's serum contained patent blue. Here, we report a case wherein patent blue affected the lipemia index. Thus, it must be noted that patent blue injection may yield inaccurate results when evaluating lipemia index. [ABSTRACT FROM AUTHOR]