1. Lactate point-of-care testing for acidosis: Cross-comparison of two devices with routine laboratory results
- Author
-
Remco van Horssen, Teska N. Schuurman, Bernadette S. Jakobs, and Monique J.M. de Groot
- Subjects
Fetal acidosis ,Point-of-care testing ,Coefficient of variation ,Clinical Biochemistry ,POCT, point-of-care testing ,Umbilical cord ,Blood gas ,TEa, Total Allowable Error ,lcsh:Chemistry ,03 medical and health sciences ,0302 clinical medicine ,medicine ,030212 general & internal medicine ,Acidosis ,lcsh:R5-920 ,030219 obstetrics & reproductive medicine ,Radiological and Ultrasound Technology ,business.industry ,ED, Emergency Department ,CV, coefficient of variation ,Routine laboratory ,Metabolic acidosis ,Medical decision making ,medicine.disease ,CCU, Critical Care Unit ,medicine.anatomical_structure ,lcsh:QD1-999 ,Anesthesia ,Lactate ,medicine.symptom ,SD, standard deviation ,business ,lcsh:Medicine (General) ,SEE, Standard Error of Estimate ,Research Article ,Biomedical engineering - Abstract
Objectives: Lactate is a major parameter in medical decision making. During labor, it is an indicator for fetal acidosis and immediate intervention. In the Emergency Department (ED), rapid analysis of lactate/blood gas is crucial for optimal patient care. Our objectives were to cross-compare-for the first time-two point-of-care testing (POCT) lactate devices with routine laboratory results using novel tight precision targets and evaluate different lactate cut-off concentrations to predict metabolic acidosis. Design and methods: Blood samples from the delivery room (n=66) and from the ED (n=85) were analyzed on two POCT devices, the StatStrip-Lactate (Nova Biomedical) and the iSTAT-1 (CG4+ cassettes, Abbott), and compared to the routine laboratory analyzer (ABL-735, Radiometer). Lactate concentrations were cross-compared between these analyzers. Results: The StatStrip correlated well with the ABL-735 (R=0.9737) and with the iSTAT-1 (R=0.9774) for lactate in umbilical cord blood. Lactate concentrations in ED samples measured on the iSTAT-1 and ABL-735 showed a correlation coefficient of R=0.9953. Analytical imprecision was excellent for lactate and pH, while for pO2 and pCO2 the coefficient of variation was relatively high using the iSTAT-1. Conclusion: Both POCT devices showed adequate analytical performance to measure lactate. The StatStrip can indicate metabolic acidosis in 1 μl blood and will be implemented at the delivery room. Keywords: Lactate, Point-of-care testing, Blood gas, Fetal acidosis
- Full Text
- View/download PDF