1. Imaging performance of single-element CdZnTe detectors for digital radiography
- Author
-
Paolo Russo, P. Chirco, M. Rossi, L. Marcello, Marcello Rossetti Conti, E. Bertolucci, E., Bertolucci, P., Chirco, M., Conti, L., Marcello, M., Rossi, Russo, Paolo, Bertolucci, Ennio, and Conti, Maurizio
- Subjects
Physics ,Nuclear and High Energy Physics ,medicine.medical_specialty ,Physics::Instrumentation and Detectors ,business.industry ,Radiography ,Resolution (electron density) ,Detector ,Biasing ,ARRAYS ,Semiconductor detector ,GAMMA-RAY DETECTORS ,Optics ,Nuclear Energy and Engineering ,Industrial radiography ,Electric field ,medicine ,Medical physics ,Electrical and Electronic Engineering ,business ,Digital radiography - Abstract
We measured the charge collection efficiency and energy resolution at 60 keV and 122 keV for single-element Cd/sub 0.8/Zn/sub 0.2/Te semiconductor detectors 2/spl times/10/spl times/10 mm/sup 3/, irradiated transverse to the electric field, in view of their possible use for digital radiography. Charge collection efficiency, energy resolution and charge collection time have been measured as a function of the bias voltage. The detector's imaging performance was evaluated, using a scanning procedure, by measurements (at 60 keV) of the line spread function and edge response function of the imaging system; metallic test objects and bone tissue were imaged. Image contrast values as low as C=10% were obtained for a minimum 5:1 S.N.R.
- Published
- 1998