1. Current phototherapy practice on Java, Indonesia
- Author
-
Mahendra T. A. Sampurna, Kinanti A. Ratnasari, Darto Saharso, Arend F. Bos, Pieter J. J. Sauer, Peter H. Dijk, and Christian V. Hulzebos
- Subjects
Curtain ,Distance ,Hyperbilirubinemia ,Irradiance ,Phototherapy ,Pediatrics ,RJ1-570 - Abstract
Abstract Background In Indonesia, the burden of severe hyperbilirubinemia is higher compared to other countries. Whether this is related to ineffective phototherapy (PT) is unknown. The aim of this study is to investigate the performance of phototherapy devices in hospitals on Java, Indonesia. Methods In 17 hospitals we measured 77 combinations of 20 different phototherapy devices, with and without curtains drawn around the incubator/crib. With a model to mimic the silhouette of an infant, we measured the irradiance levels with an Ohmeda BiliBlanket Meter II, recorded the distance between device and model, and compared these to manufacturers’ specifications. Results In nine hospitals the irradiance levels were less than required for standard PT: 50 μW/cm2/nm. Half of the distances between device and model were greater than recommended. Distance was inversely correlated with irradiance levels (R2 = 0.1838; P
- Published
- 2019
- Full Text
- View/download PDF